Digital ECG Phantom Design to Represent the Human Heart Signal for Early Test on ECG Machine in Hospital Ardila, Sella Octa; Yulianto, Endro; Sumber, Sumber
International Journal of Advanced Health Science and Technology Vol. 1 No. 1 (2021): November
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (663.633 KB) | DOI: 10.35882/ijahst.v1i1.3

Abstract

Electrocardiograph (ECG) is a diagnostic tool that can record the electrical activity of the human heart. By analyzing the resulting waveforms of the recorded electrical activity of the heart, it is possible to record and diagnose disease. Given the importance of the ECG recording device, it is necessary to check the function of the ECG recording device, namely by performing a device calibration procedure using the Phantom ECG which aims to simulate the ECG signal. The purpose of this research is to check the ECG device during repairs, besides that the Electrocardiograph (EKG) tool functions for research purposes on ECG signals or for educational purposes. Electrocardiograph (EKG) simulator or often called Phantom ECG is in principle a signal generator in the form of an ECG like signal or a recorded ECG signal. This device can be realized based on microcontroller and analog circuit. The advantage of this simulator research is that the ECG signal displayed is the original ECG recording and has an adequate ECG signal database. ECG This simulator also has the advantage of providing convenience for research on digital signal processing applications for ECG signal processing. In its application this simulator can be used as a tool to study various forms of  ECG signals. Based on the measurement results, the error value at BPM 30 and 60 is 0.00% at the sensitivity of 0.5mV, 1.0mV, and 2.0mV, then the measurement results for the error value at BPM 120 are 0.33% and at the BPM 180 value, the error value is 0.22%. From these results, it can be concluded that the highest error value is at BPM 120 with sensitivities of 0.5mV, 1.0mV, and 2.0mV.  
Nine Channel Temperature Data Logger in Measuring the Effectiveness of the Sterilization Process of Medical Instruments with Dry Sterilization Farola, Nurul Al Istigho; Gumiwang, Her; Sumber, Sumber
International Journal of Advanced Health Science and Technology Vol. 1 No. 1 (2021): November
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (283.843 KB) | DOI: 10.35882/ijahst.v1i1.5

Abstract

Measuring the temperature on the dry sterilizer is very necessary because the temperature inside the dry sterilizer has the possibility that the temperature is not the same as the temperature that has been set and is displayed on the display. If the temperature in the dry sterilizer does not match the standard setting temperature for the sterilization process, then the sterilization process is said to be imperfect The purpose of this study is to record and monitor whether the distributed temperature in the sterilization chamber corresponds to the setting temperature. The workings of the temperature data logger tool is that the type K thermocouple temperature sensor will detect the temperature which then enters the analog signal conditioning circuit which then enters the ATMegga 2560 which has been given a program and processed in such a way, then the temperature will be displayed on a 4x20 character LCD. Temperature measurement data will be saved to the SD Card every 10 seconds in the form of a TXT file. This research has been used to record 2 sterilizers and compared with the Madgetech OctTemp2000 data logger. Based on data measurements and comparisons, the average error was obtained at a temperature of 50ºC with the smallest error value of 0.7% and the largest value of 3.9%. At a temperature of 100ºC, the smallest error value is 1.6% and the largest is 10.5%. Then at a temperature of 120ºC the smallest error value is 0.0% and the largest is 8.5%. This research can be used to help analyze the distribution of temperature in a room. With these measurement results, it can be said that this study still has afairly high error value at several measurement points.
Microcontrollers Performance in Portable Electronic Stethoscope Design with a Disease Symptoms Detection Feature Hamzah, Torib Hamzah; Setioningsih, Endang Dian; Sumber, Sumber
International Journal of Advanced Health Science and Technology Vol. 1 No. 2 (2021): December
Publisher : Forum Ilmiah Teknologi dan Ilmu Kesehatan (FORITIKES)

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (1148.467 KB) | DOI: 10.35882/ijahst.v1i2.5

Abstract

One of the early examinations that is often done is to detect heart disease using a stethoscope. The electronic stethoscope consists of a membrane and tube from a conventional stethoscope coupled with a condenser microphone which is then connected to a computer. The purpose of this study is to analyze the comparison of two types of microcontrollers in the design of a portable electronic stethoscope equipped with a symptom detector. The research method used is instrumentation with 2 types of microcontrollers to design a heart sound detector. In processing the data to be displayed on the 16x2 Character LCD. Sending heart signal data for 60 seconds to produce BPM data which is processed using 2 different types of microcontrollers. The results of data collection on battery consumption of power usage on the AT mega 2560 resulted in an average saving of 0.11W. Therefore, it can be concluded that the two stethoscopes have a significant difference when compared, where the Arduino Mega 2560 is able to process data from heart signals faster than the AT mega 328P. The results of the research that have been carried out can be implemented using a system that strongly supports the needs when checking heart sound signals