Penentuan Nilai Faktor Mesin Pesawat Sinar-X Radiografi Digital Merek Shimadzu di RSUD Dr. Kanujoso Djatiwibowo Balikpapan Mamba'ul Fitriyana; Suhadi Muliyono; Kadek Subagiada
Progressive Physics Journal Vol 1 No 1 (2020): Progressive Physics Journal
Publisher : Program Studi Fisika, Jurusan Fisika, FMIPA, Universitas Mulawarman

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (596.21 KB) | DOI: 10.30872/ppj.v1i1.563

Abstract

The research on the determining machine factor (k) of Shimadzu digital radiograph X-ray machine have been done at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan. Machine factor should be required in the calculation of radiation exposure which produced by the X-ray machine. Measurement data that have been used as independent variables were current-time, focus distance to the detector and tube voltage variations of 80 kV to 100 kV, while the dependent variable was the S value. Determination of the k value was conducted by weighted linear regression between V2 and using the results of research by Seibert and Morin (2011) for the condition of calibrated X-ray machine. Thus, the reseach has obtained the machine factor of the Shimadzu digital radiograph X-ray machine at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan.