Pengaruh Slice Thickness terhadap Signal to Noise Ratio (SNR) dari Hasil Penyinaran CT Scan di RSUP Prof. Dr. I.G.N.G Ngoerah Kusumaningsih, Lusi Putu Russita; Suryatika, I Bagus Made; Trisnawati, Ni Luh Putu; Irhas, Rozi
Kappa Journal Vol 7 No 2 (2023): Agustus
Publisher : Universitas Hamzanwadi

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.29408/kpj.v7i2.7675

Abstract

A research has been carried out on the effect of slice thickness on SNR values from the results of CT Scan irradiation at Prof. Dr. I.G.N.G Ngoerah Hospital.  Variations in the thickness of the slices used are 1 mm, 3 mm, 5 mm, and 7 mm.  The tube voltage, tube current, and FOV used are constant, namely 120 kV, 200 mA, and 380 mm.  SNR value analysis was calculated using the average object value and SD background for each material.  Based on the results of a simple linear regression test, the thickness of the slices obtained has an effect of 99.9% on the SNR value.  These results indicate that the greater the thickness of the slice used, so the SNR value will increasesas the SNR value increases, it will improve image quality.
Penentuan Dosis Efektif Pada Abdo Pelvis dan Organ Kritis dari Hasil Penyinaran Computed Tomography Scanner (CT Scan) Oktar Mahardika, I Gusi Putu Gede; Suryatika, Ida Bagus Made; Putra, I Ketut; Irhas, Rozi
Kappa Journal Vol 7 No 2 (2023): Agustus
Publisher : Universitas Hamzanwadi

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.29408/kpj.v7i2.7676

Abstract

Determination of the effective dose on the abdo pelvis and organs at risk has been carried out from the results of the CT Scan irradiation at Sanglah General Hospital.The study began by collecting secondary data from the results of an abdominal CT scan, namely the CTDIvol and DLP values. Then the data was grouped based on gender of 40 patients with an age range of 20-70 years. The value of the effective dose in the abdo pelvis is obtained by multiplying the DLP value and the conversion coefficient k. While the effective dose to organs at risk is obtained from the product of the CTDIvol and wt. The results of the effective dose in the abdo pelvis were obtained in the range of 4.596-6.738 mSv for male patients while for female patients it was 4.209-5.232 mSv. The results of the effective dose in the abdo pelvis are still at the threshold set by BAPETEN, which is 13.275 mSv, so the CT scan is still suitable for use. Average effective dose values for critical organs were 0.2882 mSv for the liver, 0.8646 mSv for the kidney, and 0.5764 mSv for the gonad. The highest dose was received by the gonads. This shows that the gonads are the organs that have the highest sensitivity in receiving radiation doses in the abdo pelvis
Efek Variasi Arus Waktu Terhadap Dosis Efektif Radiasi Lensa Mata Pada Penggunaan Bismuth Shielding Pada CT-Scan Kepala Sihaloho, Monika Lestari; Sandi, Nengah; Bunawas, Bunawas; Irhas, Rozi
Kappa Journal Vol 7 No 1 (2023): April
Publisher : Universitas Hamzanwadi

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.29408/kpj.v7i1.15995

Abstract

A study has been carried out to determine the effect of time current variation on the effective dose of eye lens radiation on head CT scans. The aim of this study was to determine the effect of changing the flow of time on the effective dose of eye lens radiation and to determine the reduction in the effective dose of eye lens radiation when using bismuth shielding. The tools used are CT-scan, head mannequin, thermoliminescence dosimeter (TLD) calcium sulfate CaSO4;Dy, and bismuth shielding. The method used in this study uses the calculation of the effective dose of the 100 kV calibration curve from NuclindoLab and the equivalent dose. The measurement results show an average of each current of 31.540 mSv 100 mAs 43.947 mSv 150 mSv 59.808 msv 200 mAs 64.747 mSv 250 mAs, 137.083 mSv 300 mAs without bismuth and using bismuth 26.883 mSv 100 mAs 100 mAs, 27.847 mSv 27.84 mSv 200 mAs, 50.430 mSv 250 mAs, 81.334 mSv 300 mAs. Based on the results obtained, it is known that the effective dose obtained in the eye lens without bismuth is 67.42 ± 39.41 mSv and using bismuth is 48.85 ± 22.43 mSv with a dose reduction of 27%. As for efforts to reduce the dose received, it must be done by paying attention to the parameter settings on the CT-Scan examination such as voltage, current, slice thinkness, pitch and the use of protective equipment such as bismuth shielding.