-, Nugraha
Indonesian Institute of Sciences

Published : 1 Documents Claim Missing Document
Claim Missing Document
Check
Articles

Found 1 Documents
Search

Polyaniline as Novel Polymer Materials for Dry Electrode- Based Electrocardiography (ECG) Prima, Jeffry Omega; Pamungkas, Bayu; -, Nugraha; -, Suprijanto
Jurnal Elektronika dan Telekomunikasi Vol 18, No 1 (2018)
Publisher : Indonesian Institute of Sciences

Show Abstract | Download Original | Original Source | Check in Google Scholar | Full PDF (1202.863 KB) | DOI: 10.14203/jet.v18.1-8

Abstract

Electrocardiography (ECG) has been instrumental for early detection of cardiovascular anomalies. In this research, we successfully prototyped and evaluated the performance of a novel dry electrode as ECG sensor. Copper was selected as dry electrode material due to its good balance between conductivity and affordability. Polyaniline was used as a conductive coating to facilitate the conversion of ionic currents from the human body into electrical currents in electronic circuits. The coating was carried out via electrodeposition technique. Optimal electrodeposition time of 20 min using acetic acid as a dopant was established. This yields in the dry electrode with comparable performance to the certified wet electrode currently available in the market, shown by PQRST signal correlation between dry electrodes (this study) and the wet electrode (benchmark) which is close to zero. The dry electrode prototyped in this study is characterized to have a conductivity of 7 x 10 -4 S/mm, a resistance of 10 Ω, capacitive reactance of 140 Ω, and excellent signal stability showing a value of 256.5 μV consistently for 2 hours.