Simple and Low Cost Design of Infusion Device Analyzer Based on Arduino Nikmatul Jannah; Syaifudin Syaifudin; Liliek Soetjiatie; Muhammad Irfan Ali
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 2 No 2 (2020): August
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v2i2.4

Abstract

In the medical world, patient safety is a top priority. The number of workloads and frequency of use in the long term will affect the accuracy and precision of the equipment, therefore calibration is needed, namely the measurement activities to determine the truth of the appointment value of measuring instruments and/or measuring materials based on the standards of the Minister of Health Regulation No. 54/2015. The purpose of this study is to make the design of the Infusion Device Analyzer on flow rate parameters. The main advantage of this study is that the system can display three calibration results in one measurement at the same setting. The results of the calibration will determine the feasibility of an infusion pump or a syringe pump. This study uses the flow rate formula which is applied to the water level system to obtain the calibration results. The infrared photodiode sensor will detect the flow of water in the chamber that comes from the infusion or syringe pump. Furthermore, the sensor output will be processed by the microcontroller and the reading results are displayed on the liquid crystal display. The average measurement at a setting of 10 ml/hour is 9.36 ml/hour, at a setting of 50 ml/hour is 46.64 ml/hour and at a setting of 100 ml/hour is 96.04 ml/hour. Based on available data, this tool has an average error value of 5.69%, where the value exceeds the tolerance limit allowed by ECRI, which is ± 5%.
Design an Occlusion Calibrator using XGZP6887 and Servo Motor MG966R as a Simulator Rizki Auliya; Syaifudin Syaifudin; Liliek Soetjiatie
Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics Vol 3 No 1 (2021): February
Publisher : Department of electromedical engineering, Health Polytechnic of Surabaya, Ministry of Health Indonesia

Show Abstract | Download Original | Original Source | Check in Google Scholar | DOI: 10.35882/ijeeemi.v3i1.5

Abstract

A foreign fluid that enters the patient can cause some bodily reactions including infection, air embolism and blood clot. Side effects given will be fatal to the body, one of which occurs the blockage of the capillary vessels in the heart that can cause heart attack to stroke. The purpose of this research is to design a tool that can be used to measure maximum pressure as a form of the calibration of the syringe pump and infusion pump. The contribution of this research is that the system can simulate the presence of blockages in fluid flow and detect large pressure values detected by the Under Test Unit (UUT) with a motor peerround system that opens/closes fluid flow. Servo Motor MG966R simulate the presence of blockage with constant motor degree until the alarm UUT reads, then Sensor XGZP6887 detects the pressure generated by the blockage and processed by the microcontroller and displayed on the LCD display of the character. This study resulted in a maximum pressure average value of 7.12 Psi. The results showed that data retrieval had an error value of -0.12. This research can be implemented to perform pressure measurements on the syringe pump or infusion pump.