Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography

Muhammad Yunus (Universitas Negeri Gorontalo)
Asmiati Amir (Politeknik Kesehatan Muhammadiyah Makassar)
Mirnawati Mirnawati (Universitas Teknologi Sulawesi)
Fitriani Fitriani (Universitas Hasanuddin)
Siti Fatimah (Akademi Teknologi Industri Dewantara Palopo)



Article Info

Publish Date
30 May 2022

Abstract

The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021. This measurement uses a ruler, X-ray multimeter, and aluminum filter. The exposure process is carried out at certain kVp and mAs conditions after placing the detector on the examination table. The HVL test to determine the X-ray beam quality is carried out by placing the detector 4 cm from the edge of the chest wall on the patient's table in an upright position facing the tube. The resulting HVL values ​​are 0.37, 0.37, and 0.38. This study resulted in a p-value of 0.000 (p < 0.05). Statistically, there is a significant difference in HVL values ​​in 2019-2021. The resulting HVL value is by the tolerance limit recommended by the IAEA, which is 0.31 ≤ HVL ≤ 0.58. These values ​​indicate that mammography works well, and it is safe for the patient.

Copyrights © 2022






Journal Info

Abbrev

djtech

Publisher

Subject

Agriculture, Biological Sciences & Forestry Automotive Engineering Civil Engineering, Building, Construction & Architecture Computer Science & IT Electrical & Electronics Engineering

Description

Dewantara Journal of Technology (Dewantara J. Tech) adalah terbitan berkala yang dapat diakses secara terbuka dan didistribusikan secara nasional. Jurnal ini melingkupi berbagai topik dalam disiplin ilmu teknik, termasuk studi konseptual, studi teoritis dan eksperimental, instrumentasi, dan studi ...