Monte Carlo Simulation to Determine Deposition Ratio of Dosage in a-SI Epid with Dose Deposition on Water

Herwiningsih, Sri (Unknown)
Rianto, Sugeng (Unknown)
Yuana, Firdy (Unknown)



Article Info

Publish Date
19 May 2014

Abstract

The study aimed to explore the use of a-Si EPID as a dosimeter for IMRT delivery. The main objective of the study was to determine the ratio of dose deposited in a-Si EPID and the dose deposited in water. Linac used in the simulation was Electa Precise Linac and A-Si EPID model used in the study was based on the configuration of iView GT Elekta a-Si EPID. Study the characteristic of energy deposition at a-Si EPID and water detector shows that the energy deposition at a-Si EPID detector is higher than that of water detector at low photon energy (<0.1 MeV). This is caused by an enhancement of photoelectric interaction between photon beam and a-Si EPID detector at low energy X-rays. The simulation of dose deposition by using square fields show that the ratio between the dose deposition of a-Si EPID and the dose deposition of water detector (α) increases with an increase of the field size. The value of α ranges from 0.60 – 0.66. The similar result was shown in the simulation of dose deposition by using IMRT beam, in which the value of α ranges from 0.63 – 0.64. The α value is useful to relate the measured dose from detector to the equivalent dose of water, which is often used as a reference medium in radiotherapy dosimetry.  

Copyrights © 2014






Journal Info

Abbrev

NATURAL B

Publisher

Subject

Environmental Science

Description

Anda dapat mengakses artikel-artikel hasil penelitian khususnya bidang lingkungan dan kesehatan. Untuk informasi lebih lanjut, silakan menghubungi redaksi ...