Optimizing Analysis Of The Radiographic Image And Entrance Surface Dose Using Computed Radiography In Chest Examination

Nursama Heru Apriantoro (Politeknik Kesehatan Kemenkes Jakarta II)
Budi Santoso (Universitas Nasional)
Purwantiningsih Purwantiningsih (Universitas Nasional)
Tri Ambarsari (Universitas Nasional)



Article Info

Publish Date
01 Dec 2018

Abstract

This research has conducted about exposure factor variation analysis examination of the thorax surface dose and image noise in radiography. Do 10 (ten) times the exposure in the thorax phantom with various exposure factor such as 50 kV- 8 mAs, 50 kV-10 mAs, 55 kV- 8 mAs, 55 kV-10 mAs, 55 kV-10 mAs, 60 kV- 8 mAs, 60 kV-10 mAs, 66 kV-8 mAs, 66 kV-10 mAs, and 102 kV- 1 mAs, 102 kV - 2,5 mAs for high kV technique. For each exposure factor the pinned , TLD placed in the middle of the field to measure the radiation dose level. The optimum exposure conditions of the thorax PA examination (expressed with IgM value) occurs when using exposure factor 55 kV and 8 mAs for IgM values ​​obtained are 1.91, 6.605 for standard deviation values ​​and ESD 0.098 mGy.

Copyrights © 2018






Journal Info

Abbrev

SANITAS

Publisher

Subject

Environmental Science Health Professions Medicine & Pharmacology Public Health

Description

SANITAS: JURNAL TEKNOLOGI DAN SENI KESEHATAN (abb. SANITAS) is an open-access journal in the field of Technology and Art of Health. The journal is published twice a year in June and December. The journal aims to publish scientific research on health, case reports, literature review, to inform the ...